from ......Internal.Core import Core
from ......Internal.CommandsGroup import CommandsGroup
from ......Internal import Conversions
from ...... import enums
# noinspection PyPep8Naming,PyAttributeOutsideInit,SpellCheckingInspection
[docs]class SourceCls:
"""Source commands group definition. 6 total commands, 3 Subgroups, 1 group commands"""
def __init__(self, core: Core, parent):
self._core = core
self._cmd_group = CommandsGroup("source", core, parent)
@property
def output(self):
"""output commands group. 1 Sub-classes, 0 commands."""
if not hasattr(self, '_output'):
from .Output import OutputCls
self._output = OutputCls(self._core, self._cmd_group)
return self._output
@property
def omode(self):
"""omode commands group. 1 Sub-classes, 1 commands."""
if not hasattr(self, '_omode'):
from .Omode import OmodeCls
self._omode = OmodeCls(self._core, self._cmd_group)
return self._omode
@property
def dio(self):
"""dio commands group. 2 Sub-classes, 0 commands."""
if not hasattr(self, '_dio'):
from .Dio import DioCls
self._dio = DioCls(self._core, self._cmd_group)
return self._dio
[docs] def set(self, arg_0: enums.TriggerSource) -> None:
"""SCPI: TRIGger[:SEQuence][:IMMediate]:SOURce \n
Snippet: driver.trigger.sequence.immediate.source.set(arg_0 = enums.TriggerSource.DIO) \n
Sets or queries the trigger source. See Figure 'Overview of trigger IO system'. \n
:param arg_0: OUTPut | OMODe | DIO OUTPut Trigger source is from the output channel. OMODe Trigger source is from the different modes (CC, CR, CV, Sink, OVP, OCP, OPP and OTP) detected from the output channel. DIO Trigger source is from DIO connector at the instrument rear panel.
"""
param = Conversions.enum_scalar_to_str(arg_0, enums.TriggerSource)
self._core.io.write(f'TRIGger:SEQuence:IMMediate:SOURce {param}')
# noinspection PyTypeChecker
[docs] def get(self, arg_0: enums.TriggerSource) -> enums.TriggerSource:
"""SCPI: TRIGger[:SEQuence][:IMMediate]:SOURce \n
Snippet: value: enums.TriggerSource = driver.trigger.sequence.immediate.source.get(arg_0 = enums.TriggerSource.DIO) \n
Sets or queries the trigger source. See Figure 'Overview of trigger IO system'. \n
:param arg_0: OUTPut | OMODe | DIO OUTPut Trigger source is from the output channel. OMODe Trigger source is from the different modes (CC, CR, CV, Sink, OVP, OCP, OPP and OTP) detected from the output channel. DIO Trigger source is from DIO connector at the instrument rear panel.
:return: arg_0: OUTPut | OMODe | DIO OUTPut Trigger source is from the output channel. OMODe Trigger source is from the different modes (CC, CR, CV, Sink, OVP, OCP, OPP and OTP) detected from the output channel. DIO Trigger source is from DIO connector at the instrument rear panel."""
param = Conversions.enum_scalar_to_str(arg_0, enums.TriggerSource)
response = self._core.io.query_str(f'TRIGger:SEQuence:IMMediate:SOURce? {param}')